Mode
Continuous loading,
random access,
STAT function
Throughput
Two sampling arms
300 tests/hour for single reagent tests
190 tests/hour for two-reagents tests
Types of analysis
Endpoint, Bichromatic endpoint, Differential endpoint sample blank, Fixed time, Kinetic, Kinetic bichromatic
Samples
60 positions removable sample tray
Sample volume: 2 – 300 μL
Reagents
2 x 30 positions removable reagent trays
Refrigeration up to 12°C below room temperature
Reagent volume: 5 – 350 μL
Reaction cuvettes
80 reusable Bionex® cuvettes
Reaction volume: 210 – 350 μL
Optical system
1 halogen lamp with extended UV emission
8 interference filters at 340, 405, 505, 546, 578, 600, 650, 700 nm)
Photometric range: 0 – 3 Abs
Linearity: +/- 0,5% from 0,1 to 1,5 Abs
With a throughput up to 300 tests per hour it is the ideal solution for enhanced productivity. Wide range of validated pre-barcoded clinical chemistry and plasma and urinary proteins reagents available. 60 refrigerated positions for reagents with removable trays. 60 sample positions. Continuous loading of samples. Automatic and programmable start-up of the system. Intuitive and informative user interface allow the continuous monitoring of the system’s operations. Long walk-away capability.